                                     INSTRUMENT CONTROL PARAMETERS:    GC6850_MSD5973
                                     ------------------------------------------------

   C:\MSDCHEM\1\METHODS\Manual_170_300.M
      Wed Nov 07 21:57:47 2012

Control Information
------- -----------

Sample Inlet      :  GC
Injection Source  :  Manual
Mass Spectrometer :  Enabled


 No Sample Prep method has been assigned to this method.

=============================================================================
                                6850 GC METHOD
=============================================================================


OVEN
   Initial temp:  170 'C (On)              Maximum temp:  325 'C
   Initial time:  2.00 min                 Equilibration time:  1.00 min
   Ramps:
      #  Rate  Final temp  Final time
      1 15.00      300       40.00
      2   0.0(Off)
   Post temp:  0 'C
   Post time:  0.00 min
   Run time:  50.67 min


INLET (SPLIT/SPLITLESS)                 DETECTOR (NO DET)
   Mode:  Split                            Temperature:  280 'C (On)
   Initial temp:  280 'C (On)
   Pressure:  217.4 kPa (On)
   Split ratio:  40:1
   Split flow:  50.3 mL/min
   Total flow:  53.8 mL/min
   Gas saver:  On
   Saver flow:  20.0 mL/min
   Saver time:  2.00 min
   Gas type:  Helium


COLUMN                                  SIGNAL
   Capillary Column                        Data rate:  200 Hz
   Model Number:  J&W 122-5562UI           Type:  col comp
   DB-5MS UI                               Save Data:  Off
   Max temperature:  325 'C                Zero:  0.0 (Off)
   Nominal length:  60.0 m                 Range:  0
   Nominal diameter:  250.00 um            Fast Peaks:  Off
   Nominal film thickness:  0.25 um        Attenuation:  0
   Mode:  constant flow
   Initial flow:  1.3 mL/min
   Nominal init pressure:  217.4 kPa
   Average velocity:  30 cm/sec
   Source:  Inlet
   Outlet:  MSD
   Outlet pressure:  vacuum


COLUMN COMP
   Use:  Valve Box Heater


VALVES                                  POST RUN
                                           Post Time: 0.00 min


TIME TABLE
   Time       Specifier                     Parameter & Setpoint




                               GC Injector


     Front Injector:
        Sample Washes                 3
        Sample Pumps                  6
        Injection Volume           1.00 microliters
        Syringe Size               10.0 microliters
        PreInj Solvent A Washes       3
        PreInj Solvent B Washes       3
        PostInj Solvent A Washes      3
        PostInj Solvent B Washes      3
        Viscosity Delay               0 seconds
        Plunger Speed              Fast
        PreInjection Dwell         0.00 minutes
        PostInjection Dwell        0.00 minutes


     Back Injector:
No parameters specified

 Column 1 Inventory Number : AB001
                                MS ACQUISITION PARAMETERS


General Information
------- -----------

Tune File                : atune.u
Acquistion Mode          : Scan


MS Information
-- -----------

Solvent Delay            : 3.90 min

EMV Mode                 : Relative
Relative Voltage         : 0
Resulting EM Voltage     : 1812

[Scan Parameters]

Low Mass                 : 29.0
High Mass                : 550.0
Threshold                : 150
Sample #                 : 2       A/D Samples    4

[MSZones]

MS Source                : 230 C   maximum 250 C
MS Quad                  : 150 C   maximum 200 C


                             END OF MS ACQUISITION PARAMETERS


                              TUNE PARAMETERS for SN: us53941194
                        -----------------------------

 Trace Ion Detection is OFF.

 EMISSION    :      34.610
 ENERGY      :      69.922
 REPELLER    :      26.772
 IONFOCUS    :      79.718
 ENTRANCE_LE :       9.500
 EMVOLTS     :    1811.765
                               Actual EMV  :    1811.77
                               GAIN FACTOR :       0.34
 AMUGAIN     :    2355.000
 AMUOFFSET   :     124.000
 FILAMENT    :       1.000
 DCPOLARITY  :       0.000
 ENTLENSOFFS :      18.322
 MASSGAIN    :      80.000   
 MASSOFFSET  :     -13.000   

                           END OF TUNE PARAMETERS
                      ------------------------------------



                                           END OF INSTRUMENT CONTROL PARAMETERS
                                           ------------------------------------